The F2A3X Frequency to Analog Converter is a DIN rail module that converts a frequency input signal into a proportional analog voltage (0-5 Vdc) or current (4-20 mA) output. The output signal is electrically isolated from the input signal and input power source, effectively eliminating troublesome ground loops. The input signal can be supplied from a Monarch sensor (measuring RPM for example) or any source of digital signal not exceeding 12 volts. The F2A3X is completely user-programmable using the free downloadable PM Remote Software. A NIST Traceable Certificate is included.
The F2A3X Frequency to Analog Converter is a DIN rail module that converts a frequency input signal into a proportional analog voltage (0-5 Vdc) or current (4-20 mA) output. The output signal is electrically isolated from the input signal and input power source, effectively eliminating troublesome ground loops. The input signal can be supplied from a Monarch sensor (measuring RPM for example) or any source of digital signal not exceeding 12 volts. The F2A3X is completely user-programmable using the free downloadable PM Remote Software. A NIST Traceable Certificate is included.
Specification | Specification value |
---|---|
Input Range | 5-999,990 RPM speeds below 5 RPM possible with multiple pulses per revolution (0.083 Hz to 15.0 kHz) |
Input Configuration and Voltage Range | 1 to 9,999 pulse per revolution or use a scale factor of 0.0001 to 99,999; PC software programmable, TTL input and 1.1 V to 25 Vdc signals – Internal Jumper for ±1 to ±25 V ac |
Analog Output | Voltage: 0-5 Vdc, 5 mA max load, Isolated or 4-20 mA Isolated, 500 ohms max load, Internal 12 V compliance voltage, 16-bit resolution; full scale and offset RPM ranges PC programmable |
Accuracy/Resolution | 0.005% of full scale output / 76 µvolts or 30.5 nanoamps |
Output Update | Software selectable up to 244 times/sec; dependent on the input frequency |
Memory | Max and min. recall via PC software |
Dimensions | 1/8 DIN by 3.94 in. [100 mm] deep |
Input Power | Standard 12-24 Vdc, 4.5 W max or optional 12 Vdc to 24 Vdc isolated 4.5 W max |
Sensor Excitation | 10 V dc @ 60 mA standard or 5 Vdc @ 60 mA optional |
Pulse Repeater Output | 0-5 V TTL compatible, one pulse out for each pulse in; polarity is software selectable |
Communications | Ethernet RJ45 or USB type B |
Alarm Capability | Optional alarm with relay output Setpoints: high or low alarm limit, latching or non-latching PC programmable |
Alarm Outputs | 1 form C relay contact, rated 1A at 115 or 230 Vac |
Alarm Reset | Automatic or manual reset; front panel push button or remote reset via PM Remote Software |
Environmental | Indoor use only; installation category II per IEC 664 Temperature: -10 °C to 50 °C, operating per IEC 61010-1 Humidity: 80% max to temps up to 31 °C, decreasing linearity to 50% RH at 40 °C |
Electrical Safety | Meets EN61010-1:2001, EC low voltage directive 2006/95/EC |
PM Remote Software is a free downloadable Windows-based software application that allows users to quickly and easily customize the configuration of Monarch Instrument’s ACT panel tachometers and F2A1X and F2A3X Frequency to Analog Modules. Set the mode of operation to RPM, RPS, or Frequency and select the input scale (pulses per revolution). Real-time data can be displayed directly on the PC along with min and max values. Decimal places and display update rate are user configurable.
Features
Harmonized Tariff Code: 9029205000
Country of Origin: US
Export Control Classification Number (ECCN): EAR99