This battery-powered, stand-alone device with display can record up to 130,000 samples for logging, redundant data backup or archive. It is available in a wide variety of ranges.
Includes: USB cable and Track-It Software download
* PURCHASE DIRECT FROM MANUFACTURER
* FREE LIFETIME TECH SUPPORT
* FREE SHIPPING
* TAX FREE
- Description
- Features
- Specifications
- Track-it™ Software
- Android App
- Shipping Codes
- Downloads
- Application Notes
The Track-It™ Pressure Transmitter/Data Loggers with Digital Display are 24Vdc powered 4 wire transmitters that can also record up to 130,000 samples for logging, redundant data backup, or archive. They can be configured to record both pressure or vacuum and ambient temperature or just pressure or vacuum alone to maximize data storage space. Real-time data, alarms, and min./max information can be displayed on the multiline LCD. The unit is easily configured using the free downloadable Track-It™ Software; simply connect the included USB cable to the logger and an open USB port on your PC and the Track-It™ Software automatically identifies the logger. Default configuration of the unit is 4mA = 0 and 20mA = full scale. Custom scaling over a smaller range can be accommodated using Track-It™ Software. The data logging sample storage rate can be set from 1 sample every 2 seconds up to 1 sample every 24 hours. The onboard data storage is non-volatile so data will not be lost in the event of a power loss. Track-It™ Pressure Transmitter/Data Loggers can be ordered in various ranges. The logger is packaged in a rugged anodized aluminum and stainless steel housing and has a standard 1/4" NPT fitting (1/2" NPT for 8700 and 14,500 ranges).
The Track-It™ Pressure Transmitter/Data Loggers are different than the Pressure Display Stand-alone Data Loggers which do not transmit but store data internally and download through a USB cable to a PC.
Typical Uses
- Industrial Process
- Food Manufacturing
- Water Supply
- Hydrant Water Pressure
- Building Supply
- Municipalities
- HVAC
- Oil and Gas
- Petro Chemical
- Compressors
- Pumps
- Hydrostatic Testing
- Water and Wastewater
- Production Lines
- Hydraulics
Which one is right for me?
Features
- Isolated 4-20mA output prevents ground loops
- Display live and min./max readings
- Alarm setpoints and indicators
- User-definable engineering units
- USB interface (cable included)
- Powerful interactive Track-It Software package included
- Rugged IP65 package
- Temperature range: –4 to 185°F (-20 to 85°C )
- Available ranges: 3, 35, 150, 350, 550, 2000, 5800, 8700 and 14,500 PSI
- Available Torr ranges: 0 to -380, 0 to -760, 760 to 0, and 760 to -760
- 150% over-range protection
- ASME Grade 3A Accuracy
- ¼” NPT male connection (1/2" NPT male for 8700 and 14,500 ranges)
- High accuracy and repeatability
- Accurate internal clock (1 minute/year)
- NIST Calibration Certificate available
- Swivel head 300° and rotate display 180°
Specifications*
General
Record |
Sample rates: user configured 1 every 2 seconds to 1 every 24 hours Number of samples: 130,000 (setup dependent) |
Display |
Two-line, 6-digit 7-segment LCD And 5-digit alphanumeric display with custom icons, multi-segment bar graph for 0-100% indication, backlit |
Record Trigger |
Two independent triggers Multiple trigger modes: instantaneous, button control, on alarm, time and date (start and stop), day of week |
Record Mode |
Fill to end of memory; cyclic, number of samples, time duration |
Alarms |
2 user-programmable alarms (high or low) |
Communication |
Mini USB connection |
Software |
Track-It™ Software — program device, view data (historic or real-time), export to Excel; simple and advanced modes |
Enclosure |
Material: anodized aluminum And 316L stainless steel Dimensions: 2.94” dia. X 1.82” deep X 4.78” tall (including port threads) |
Measurement
Pressure Ranges |
0-3, 0-35, 0-150, 0-350, 0-550, 0-2000, 0-5800, 0-8700, or 0-14,500 PSI |
Vacuum Ranges |
0 to -380, 0 to -760, 760 to 0, and 760 to -760 Torr |
Accuracy |
± 0.25% |
Resolution |
0.01% |
Overpressure Rating |
1.5 times max. |
Port Connection | 1/4” NPT male (1/2" NPT male for 8700 and 14,500 ranges) |
Temperature Range | -20 To +85°C / -4 To 185°F |
Resolution | 0.1°C / 0.1°F |
Transmitter
Output | Isolated 4-20mA, +/-0.06% F.S. |
Resolution | 6 UA |
Transmitter Input Power | 24 Vdc +/- 10% |
Isolation | 500 V To 24 Vdc supply |
Update Rate | 125 msec |
*Specifications are subject to change without notice.
Track-It™ Software
Track-It™ Software is a powerful Windows-based software package that is included with every Track-It data logger. It allows for easy setup, retrieval, interpretation, and export of the recorded data.
Point and click to select your sample interval, alarm setpoints, engineering units, start recording triggers, and LCDs.
View a numerical data table or graphical representation of data readings relative to time. Use the such as zoom, cursor, min. and max to examine and analyze the record. Track-It Software turns your PC into a real-time data acquisition system by allowing you to stream graphical or tabular data directly to your PC (capability varies by model). Use the one-click export function to automatically export stored data into an Excel spreadsheet.
Compatible with Win XP/Vista/7/8/10, 32 and 64 Bit
Track-It™ Transport Android App
The Track-It Pressure Transmitter Data Logger is compatible with the free Track-It Transport Android application. The Transport Application allows you to use your Android device and an On-The-Go (OTG) cable to retrieve data from the logger while in the field. Record start/stop and erase memory is also available.
The Track-It Transport App is available for free on Google Play:
Shipping Codes for the Track-It Pressure Transmitter/Data Logger with Digital Display (all models):
Harmonized Tariff Code: 9026200000
Country of Origin: US
Export Control Classification Number (ECCN): EAR99
Literature
Software
Knowledgebase Documents
Pre-installation Checklist and Installation of Track-It Drivers and Software
Track-It Frequently Asked Questions
Application Notes
Hydrostatic Pressure Testing Requirements for Water Distribution System Installation or Replacement Contractors are required to either test or hire certified testing companies to perform pressure and leakage testing in accordance with the applicable AWWA Standards and Specifications. The testing company is required to submit certified leakage testing results to the system operators or owners READ MORE... |
Data Loggers - Which one is right for me? Display Logger Gauge? Transmitter or internal data logging? When do I need it to be inconspicuous? |